Health Canada published a recall notice on Nov. 25 for three models of the HeartSine Samaritan PAD automated external defibrillator, or AED. Stryker, the company behind the notice, has identified a potential problem with the electrical connection between the device and its battery and pad cartridge.
The notice lists three models:
For each model, the notice says there are more than 10 lot or serial numbers and tells readers to contact the manufacturer for them. The official notice does not list the numbers itself.
According to the notice, the Pad-Pak or Pediatric-Pak connects to the AED through spring-loaded electrical contacts, also known as pogo pins. Stryker says oxidation of a pogo pin may increase electrical resistance over time and weaken the internal spring.
If that condition progresses, the notice says, the electrical connection between the Pad-Pak or Pediatric-Pak battery and the AED may become intermittent or may be lost. The notice does not report any injuries or incidents.
The notice says only that anyone who needs more information should contact the manufacturer. The manufacturer is listed as Heartsine Technologies, Ltd., at 207 Airport Road West, Belfast, Northern Ireland, United Kingdom, BT3 9ED. The notice does not give a remedy, a sale period or a phone number.
Anyone who owns or manages a HeartSine Samaritan PAD 350P, 360P or 500P can check the model number against the three listed above. Because the lot and serial numbers are not published in the notice, the manufacturer is the source for which individual units are included.